Dose reduction in mammography with photon counting imagingShow others and affiliations
2004 (English)In: Proceedings of SPIE - The International Society for Optical Engineering, 2004, no 1, p. 457-465Conference paper, Published paper (Refereed)
Abstract [en]
The purpose of this study was to investigate if the glandular dose to the breast in mammography can significantly be reduced without compromising image quality, when using photon counting technology, in a multi-slit scanning photon counting detector, compared to a conventional film mammography system and commercial available digital mammography systems with TFT-array detectors. A CDMAM phantom study, with two different thicknesses of additional PMMA absorber, 4 cm and 7 cm respectively, has shown that multi-slit scanning photon counting detector technology can reduce the dose, without reducing the image quality. This comparison was made to two commercial available digital mammography systems Senographe 2000D (from GEMS) and Selenia (from Lorad). The results show that dose can be reduced with 63% to 77%, depending on object thickness, when using XCT for mammography. This dose reduction has also been verified clinically through a small pilot study with patients and specimen, where the comparison was made between XCT and film.
Place, publisher, year, edition, pages
2004. no 1, p. 457-465
Series
Proceedings of SPIE - The International Society for Optical Engineering, ISSN 0277-786X ; 5368
Keywords [en]
CDMAM phantom, Digital mammography, Dose reduction, Photon counting, Arrays, Gold, Image quality, Oncology, Polymethyl methacrylates, Radiation detectors, Thickness measurement, X ray analysis, CDMAM phantoms, Mammography
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
URN: urn:nbn:se:kth:diva-157433DOI: 10.1117/12.530649ISI: 000222012700050Scopus ID: 2-s2.0-3042546832OAI: oai:DiVA.org:kth-157433DiVA, id: diva2:771175
Conference
Medical Imaging 2004: Physics of Medical Imaging, 15 February 2004 through 17 February 2004, San Diego, CA
Note
QC 20141212
2014-12-122014-12-092022-06-23Bibliographically approved