Scatter rejection in scanned multi-slit digital mammography
2004 (English)In: Proceedings of SPIE - The International Society for Optical Engineering, 2004, no 1, p. 478-487Conference paper, Published paper (Refereed)
Abstract [en]
Measurements and Monte Carlo simulations were used to investigate the scatter properties of a scanned multi-slit digital mammography system. Scatter to primary ratio (S/P) in the center of the image field was calculated for different thickness of breast equivalent material and different tube potentials. The simulated model also varied the angular acceptance, the number of slits and the distance between the slits of a dedicated scatter rejection device. In addition to the expected scatter from the breast equivalent material, scatter within the detector contributes to the S/P-ratio. The main part of the scatter is identified as coming from this process. Measured total 5/P-ratios below 3% are reported for breast range 3-8 cm. The scatter-DQE is used as figure-of-merit for comparison to other imaging geometries and scatter rejection schemes.
Place, publisher, year, edition, pages
2004. no 1, p. 478-487
Series
Proceedings of SPIE - The International Society for Optical Engineering, ISSN 0277-786X ; 5368
Keywords [en]
Digital, Mammography, Rejection, Scanning, Scatter, Slit, Absorption, Computer simulation, Image quality, Mathematical models, Monte Carlo methods, Scattering, Signal to noise ratio, Tumors, Digital scatter
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
URN: urn:nbn:se:kth:diva-157435DOI: 10.1117/12.535382ISI: 000222012700052Scopus ID: 2-s2.0-3042641613OAI: oai:DiVA.org:kth-157435DiVA, id: diva2:771154
Conference
Medical Imaging 2004: Physics of Medical Imaging, 15 February 2004 through 17 February 2004, San Diego, CA
Note
QC 20141212
2014-12-122014-12-092022-06-23Bibliographically approved