Multimodal Brain Tumor Tissue Identification: Integration of Optical Guidance and Quantitative MRI in Neuronavigated Biopsies
Linköping University, Department of Biomedical Engineering, Division of Biomedical Engineering. Linköping University, Faculty of Science & Engineering.ORCID iD: 0000-0002-7987-9841
2025 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

In neurosurgery, it is difficult to safely and reliably sample brain tumor tissue. Within the framework of this thesis, a method for optical guidance during navigated biopsy procedures was introduced. Furthermore, a workflow for multimodal analysis of tumor tissue was developed, comparing optical signals with intensities from magnetic resonance imaging (MRI) and neuropathological parameters. The five studies included in this thesis were carried out in close collaboration with the Department of Neurosurgery at Linköping University Hospital.

In Paper I, the optical technology is presented and experimentally evaluated. The system measures three signals: fluorescence, indicating high-grade tumor tissue; microcirculation, which detects blood vessels; and tissue light intensity, providing criteria for the type of tissue, such as gray or white matter. The parameters are displayed in real-time during surgery, where tumor tissue is identified as a fluorescence peak, and high, pulsatile blood flow could indicate vascular structures. In Paper II, the optical system was integrated into the frameless neuronavigation system and evaluated in a phantom before clinical measurements in three biopsy patients. Based on the integration, a multimodal workflow was introduced, including data processing, and registration to the neuronavigation coordinate system. In the clinical setting, good ability to identify tumor tissue and blood flow changes was confirmed.

Paper III presents a multiparametric analysis of data from 20 navigated biopsies where the optical technique was used. In each biopsy procedure, the biopsy needle was inserted into the brain in a stepwise manner with the optical probe inside. At each step, the optical signals were measured, the measurement position identified through the neuronavigation system, and direct feedback on tissue characteristics was received. The optical system offers an indication of where a diagnostic specimen can be sampled. Fluorescence peaks were found in high-grade tumors such as glioblastoma, astrocytoma, and lymphoma. Compared to the routine biopsy procedure, the procedure time could be shortened if fluorescence peaks are found.

To gain greater knowledge of the infiltrative zone between non-tumor and tumor tissue, the multimodal workflow was extended with two quantitative MRI (QMRI) methods: relaxometry (Paper IV) and diffusion MRI (dMRI) (Paper V). Relaxometry analyzes the biophysical basis of tissue relaxivity, e.g., before and after administration of a contrast medium. Along the biopsy trajectory, the difference in relaxivity was compared with the presence of fluorescence peaks, radiological definition of tumor, and neuropathological parameters. Increased relaxivity after contrast administration was not tumor-specific, however, the largest increases were found in tumor tissue. Diffusion MRI, or the study of the movement of water molecules in tissue, was compared in the biopsied volume and the contralateral normal-appearing white matter. Trends of lower anisotropy, kurtosis, and variance as well as increased diffusivity were found in patients with glioblastoma and astrocytoma. Furthermore, a subset of diffusion scalar maps was identified for future studies towards the clinical utility of dMRI for brain tumor patients.

In summary, probe-based optical guidance and a workflow for multimodal analysis, including QMRI data, was developed and integrated into the frameless brain tumor biopsy procedure. The techniques were combined for analysis with conventional MRI and neuropathological parameters. Quantitative MRI and optical guidance can provide further insights into tumor tissue identification and may, in extension, result in better patient care.

Abstract [sv]

Inom neurokirurgi är det svårt att ta vävnadsprover från hjärntumörer på ett säkert och tillförlitligt sätt. Inom ramen för detta avhandlingsarbete introducerades en metod för optisk guidning under kirurgin när ett vävnadsprov tas, en så kallad biopsi. Vidare utvecklades ett arbetsflöde för multimodal analys av tumörvävnad, där optiska signaler jämfördes med intensiteter från magnetisk resonanstomografi (MRT) och neuropatologiska parametrar vilka ligger till grund för diagnos. De fem studier som ingår i avhandlingen har genomförts i nära samarbete med Neurokirurgiska kliniken på Linköpings universitetssjukhus.

I delarbete I presenteras och utvärderas den optiska tekniken i laboratoriemiljö och på vävnadspreparat. Systemet mäter tre signaler: fluorescens, vilket indikerar höggradig tumörvävnad; perfusion, som kan påvisa blodkärl; och vävnadens ljusintensitet, vilket indikerar vävnadens typ, t.ex. grå eller vit substans. Parametrarna visas i realtid under operationen, där tumörvävnad identifieras av en fluorescenstopp, och en högt, pulserande perfusionssignal indikerar vaskulära strukturer. I delarbete II integrerades det optiska systemet i det ramlösa neuronavigationssystemet och utvärderades i en fantom före kliniska mätningar i tre biopsipatienter. Baserat på integrationen introducerades ett multimodalt arbetsflöde, inklusive databehandling och registrering till neuronavigationssystemets koordinatsystem. De kliniska mätningarna bekräftade god förmåga att identifiera både tumörvävnad och blodflödesförändringar.

I delarbete III presenteras en multiparametrisk analys av data från 20 neurokirurgiska tumörbiopsier där den optiska tekniken användes. Vid varje operation fördes biopsinålen stegvis in i hjärnan med den fiber-optiska proben inuti. I varje steg mättes de optiska signalerna, mätpositionen identifierades genom neuronavigationssystemet och kirurgen fick direkt återkoppling om vävnadens egenskaper. Det optiska systemet ger en indikation på var ett diagnostiskt vävnadsprov kan tas. Fluorescenstoppar hittades i höggradiga tumörer: glioblastom, astrocytom och lymfom. Jämfört med den rutinmässiga biopsiproceduren skulle operationstiden kunna förkortas om fluorescenstoppar upptäcks.

För att få ökad kunskap om den diffusa gränsen mellan frisk vävnad och tumörvävnad utökades det multimodala arbetsflödet med två kvantitativa MRI-metoder (QMRI): relaxometri (delarbete IV) och diffusions-MRI (dMRI) (delarbete V). Relaxometri analyserar den biofysiska grunden för vävnadens återgång till jämviktsläge efter att energi tillförts, där skillnader i relaxometri, t.ex. före och efter administrering av ett kontrastmedel, kan påvisa tumörvävnad. Längs biopsibanan jämfördes skillnaden i relaxivitet med förekomst av fluorescenstoppar, radiologisk definition av tumör och neuropatologiska parametrar. Ökad relaxivitet efter kontrastinjektion var inte specifik för tumörer, men de största ökningarna sågs i tumörvävnad. Diffusionsviktad MRT (dMRI), eller studiet av vattenmolekylers rö-relse i vävnad, jämfördes i biopsivolymen och den kontralaterala, till synes normala, vita vävnaden. Trender av lägre anisotropi, kurtosis och varians samt ökad diffusivitet sågs hos patienter med glioblastom och astrocytom. Som ett steg mot klinisk användbarhet identifierades ett urval av skalärkartor för framtida studier.

Sammanfattningsvis har ett system för optisk navigering i realtid utvecklats och integrerats vid hjärntumörsbiopsier. Vidare har ett kombinerat arbetsflöde för gemensam analys av optiska och kvantitativa MRT-parametrar tillsammans med konventionell MRT och neuropatologiska parametrar utvecklats. Kvantitativa MRT-metoder och optisk guidning kan ge nya insikter om höggradig tumörvävnad och på sikt bidra till en bättre patientvård.

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press, 2025. , p. 100
Series
Linköping Studies in Science and Technology. Dissertations, ISSN 0345-7524 ; 2440
Keywords [en]
Fluorescence, Spectroscopy, Glioblastoma, Laser Doppler flowmetry, Microstructure imaging, MRI relaxometry
National Category
Radiology and Medical Imaging
Identifiers
URN: urn:nbn:se:liu:diva-217376DOI: 10.3384/9789181180244ISBN: 9789181180237 (print)ISBN: 9789181180244 (electronic)OAI: oai:DiVA.org:liu-217376DiVA, id: diva2:1994645
Public defence
2025-10-03, Hugo Theorell, building 440, Campus US, Linköping, 09:00 (Swedish)
Opponent
Supervisors
Available from: 2025-09-03 Created: 2025-09-03 Last updated: 2025-09-03Bibliographically approved
List of papers
1. FluoRa - a System for Combined Fluorescence and Microcirculation Measurements in Brain Tumor Surgery
Open this publication in new window or tab >>FluoRa - a System for Combined Fluorescence and Microcirculation Measurements in Brain Tumor Surgery
Show others...
2021 (English)In: 2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), IEEE , 2021, p. 1512-1515Conference paper, Published paper (Refereed)
Abstract [en]

In brain tumor surgery it is difficult to distinguish the marginal zone with the naked eye. Fluorescence techniques can help identifying tumor tissue in the zone during resection and biopsy procedures. In this paper a novel system for combined real-time measurements of PpIX-fluorescence, microcirculation and tissue grey-whiteness is presented and experimentally evaluated. The system consists of a fluorescence hardware with a sensitive CCD spectrometer for PpIX peak detection, a laser Doppler system, optical probes, and a LabView software. System evaluation was done on static fluorescing material, human skin, and brain tumor tissue. The static material indicates reproducibility, the skin measurements exemplify simultaneous fluorescence and microcirculation measurement in real-time, and the tumor tissue showed PpIX peaks. These decreased over time, as expected, due to photo bleaching. In addition, the system was prepared for clinical use and thus laser- and electrical safety issues were considered. In summary, a system for multiparameter measurements during neurosurgery was successfully evaluated in an experimental environment. As a next step the system will be applied in clinical brain tumor biopsies and resections.

Place, publisher, year, edition, pages
IEEE, 2021
Series
IEEE Engineering in Medicine and Biology Society Conference Proceedings, ISSN 1557-170X
National Category
Medical Laboratory Technologies
Identifiers
urn:nbn:se:liu:diva-184568 (URN)10.1109/EMBC46164.2021.9629801 (DOI)000760910501122 ()34891572 (PubMedID)9781728111797 (ISBN)
Conference
43rd Annual International Conference of the IEEE-Engineering-in-Medicine-and-Biology-Society (IEEE EMBC), ELECTR NETWORK, nov 01-05, 2021
Note

Funding Agencies|Swedish Foundation for Strategic ResearchSwedish Foundation for Strategic Research [RMX18-0056]

Available from: 2022-05-02 Created: 2022-05-02 Last updated: 2025-09-03
2. Combined Use of Frameless Neuronavigation and In Situ Optical Guidance in Brain Tumor Needle Biopsies
Open this publication in new window or tab >>Combined Use of Frameless Neuronavigation and In Situ Optical Guidance in Brain Tumor Needle Biopsies
2023 (English)In: Brain Sciences, E-ISSN 2076-3425, Vol. 13, no 5, article id 809Article in journal (Refereed) Published
Abstract [en]

Brain tumor needle biopsies are performed to retrieve tissue samples for neuropathological analysis. Although preoperative images guide the procedure, there are risks of hemorrhage and sampling of non-tumor tissue. This study aimed to develop and evaluate a method for frameless one-insertion needle biopsies with in situ optical guidance and present a processing pipeline for combined postoperative analysis of optical, MRI, and neuropathological data. An optical system for quantified feedback on tissue microcirculation, gray-whiteness, and the presence of a tumor (protoporphyrin IX (PpIX) accumulation) with a one-insertion optical probe was integrated into a needle biopsy kit that was used for frameless neuronavigation. In Python, a pipeline for signal processing, image registration, and coordinate transformation was set up. The Euclidian distances between the pre- and postoperative coordinates were calculated. The proposed workflow was evaluated on static references, a phantom, and three patients with suspected high-grade gliomas. In total, six biopsy samples that overlapped with the region of the highest PpIX peak without increased microcirculation were taken. The samples were confirmed as being tumorous and postoperative imaging was used to define the biopsy locations. A 2.5 ± 1.2 mm difference between the pre- and postoperative coordinates was found. Optical guidance in frameless brain tumor biopsies could offer benefits such as quantified in situ indication of high-grade tumor tissue and indications of increased blood flow along the needle trajectory before the tissue is removed. Additionally, postoperative visualization enables the combined analysis of MRI, optical, and neuropathological data.

Keywords
MRI, five-aminolevulinic acid (5-ALA), image-guidance, laser Doppler flowmetry (LDF), microcirculation, neurosurgery
National Category
Neurology Medical Engineering
Identifiers
urn:nbn:se:liu:diva-194538 (URN)10.3390/brainsci13050809 (DOI)000995535200001 ()37239281 (PubMedID)2-s2.0-85160276818 (Scopus ID)
Note

Funding: Swedish Foundation for Strategic Research [RMX18-0056]

Available from: 2023-06-13 Created: 2023-06-13 Last updated: 2025-09-03
3. In situ optical feedback in brain tumor biopsy: A multiparametric analysis
Open this publication in new window or tab >>In situ optical feedback in brain tumor biopsy: A multiparametric analysis
Show others...
2024 (English)In: Neuro-Oncology Advances, E-ISSN 2632-2498, Vol. 6, no 1, article id vdae175Article in journal (Refereed) Published
Abstract [en]

Background Brain tumor needle biopsy interventions are inflicted with nondiagnostic or biased sampling in up to 25% and hemorrhage, including asymptomatic cases, in up to 60%. To identify diagnostic tissue and sites with increased microcirculation, intraoperative optical techniques have been suggested. The aim of this study was to investigate the clinical implications of in situ optical guidance in frameless navigated tumor biopsies.Methods Real-time feedback on protoporphyrin IX (PpIX) fluorescence, microcirculation, and gray-whiteness was given before tissue sampling (272 positions) in 20 patients along 21 trajectories in total. The primary variables of investigation were fluorescence in relation to neuropathological findings and gadolinium (Gd) enhancement, increased cerebral microcirculation in relation to bleeding incidence, number of trajectories, and impact on operation time.Results PpIX fluorescence was detected in Glioblastoma IDH-wildtype CNS WHO grade 4 (n = 12), Primary diffuse large B-cell lymphoma (n = 3), astrocytoma IDH-mutated CNS WHO grade 4 (n = 1) (Ki67 indices >= 15%). For 2 patients, no PpIX fluorescence or Gd was found, although samples contained tumorous tissue (Ki67 index 6%). Increased microcirculation was found along 9 trajectories (34 sites), located in cortical, tumorous, or tentorium regions. Postoperative bleedings (n = 10, nine asymptomatic) were related to skull opening or tissue sampling. This study strengthens the proposed independence from intraoperative neuropathology as PpIX fluorescence is detected. Objective real-time feedback resulted in fewer trajectories compared to previous studies indicating reduced operation time.Conclusions The integrated optical guidance system provides real-time feedback in situ, increasing certainty and precision of diagnostic tissue before sampling during frameless brain tumor biopsies. Taking samples from brain tumors can be risky, and sometimes patients do not receive a complete diagnosis because surgeons may have not sampled the right part of the tumor. The authors of this study wanted to see if they could use tools that provide real-time feedback during biopsies to help surgeons locate the abnormal areas better. To do this, they used a tool that analyzes the reaction of light with different tissues, and they tested this on 20 patients. The results showed that this tool helped identify the right areas of the tumor to sample and reduced the number of times a biopsy was attempted.

Place, publisher, year, edition, pages
OXFORD UNIV PRESS, 2024
Keywords
5-aminolevulinic acid; digital pathology; glioblastoma; microcirculation; neuronavigation
National Category
Other Medical Engineering
Identifiers
urn:nbn:se:liu:diva-210707 (URN)10.1093/noajnl/vdae175 (DOI)001379170200001 ()39687792 (PubMedID)2-s2.0-85212764835 (Scopus ID)
Note

Funding Agencies|Swedish Foundation for Strategic Research [RMX18-0056]

Available from: 2025-01-10 Created: 2025-01-10 Last updated: 2025-09-03
4. Quantitative MRI relaxometry in brain tumor needle biopsies: Multimodal comparison with tissue fluorescence, radiology, and neuropathology
Open this publication in new window or tab >>Quantitative MRI relaxometry in brain tumor needle biopsies: Multimodal comparison with tissue fluorescence, radiology, and neuropathology
Show others...
2025 (English)In: PLOS ONE, E-ISSN 1932-6203, Vol. 20, no 7, article id e0326765Article in journal (Refereed) Published
Abstract [en]

Background Quantitative MRI (qMRI) relaxometry holds potential for brain tumor identification beyond contrast enhancement on conventional images. However, clinical implementation is limited by long acquisition times, changing conditions between imaging and surgery, and lack of correlation with standard techniques.Purpose To extend a methodology for multimodal data analysis to relaxometry data. To integrate relaxometry into the burr hole needle biopsy procedure with optical guidance, setup a workflow for multimodal data processing and analysis, and apply the methodology in a clinical setting.Methods Multi-dimensional multi-echo relaxometry data (2x6 min) was acquired in addition to the clinical imaging protocol. Relaxation rate and proton density maps, as well as their differences were calculated before (R1, R2) and after gadolinium contrast-agent administration (R1Gd, R2Gd). Radiological volumes of interest (VOIs: tumor, edema, white matter, and biopsy) were defined on clinical images. Rate distribution changes were analyzed on three levels: the biopsied volume, along the needle trajectory (4x4x4 mm3 volumes), and VOIs. Increased R1Gd and R2Gd were compared to indications from 5-aminolevulinic acid-induced fluorescence and detailed neuropathological evaluation.Results Neuropathological analysis confirmed seven glioblastoma, one lymphoma, and one non-tumorous diagnosis. Increased R1Gd was found in all biopsied volumes, although tumorous volumes presented larger R1Gd increase (3-9 times) compared to volumes dominated by necrotic or non-tumorous tissue. Along the trajectory, increased R1Gd and R2Gd were not tumor-specific, however, the greatest R1Gd shifts were found in or adjacent to radiologically defined tumorous tissue. Increased relaxation rates corresponded to 82% and 45% (R1Gd: phi = 0.35, R2Gd: phi = 0.27) of fluorescence peaks. In the radiological VOIs, increased R1Gd and R2Gd were found in tumorous tissue, a slight right shift in edematous tissue, and negligible changes in white matter.Conclusion Combined analysis suggests increased R1Gd together with fluorescence peaks as a marker for tumor tissue. The presented multimodal approach provides a workflow toward clinical translation of relaxometry.

Place, publisher, year, edition, pages
PUBLIC LIBRARY SCIENCE, 2025
National Category
Radiology and Medical Imaging
Identifiers
urn:nbn:se:liu:diva-216906 (URN)10.1371/journal.pone.0326765 (DOI)001524179100016 ()40623087 (PubMedID)2-s2.0-105010029011 (Scopus ID)
Note

Funding Agencies|Swedish Foundation for Strategic Research [RMX18-0056]; Swedish Agency for Innovation Systems (Vinnova) [2023-03943]; Vinnova [2023-03943] Funding Source: Vinnova; Swedish Foundation for Strategic Research (SSF) [RMX18-0056] Funding Source: Swedish Foundation for Strategic Research (SSF); Swedish Research Council [2023-03943] Funding Source: Swedish Research Council

Available from: 2025-08-28 Created: 2025-08-28 Last updated: 2025-10-07

Open Access in DiVA

fulltext(22165 kB)599 downloads
File information
File name FULLTEXT01.pdfFile size 22165 kBChecksum SHA-512
21233d46fbb04c3a32a7e46fcac866f95d9176834d608a252591c10d99be94bb8134f161d27b25020834adda5943972c6b37dde76500e6bf1e47a6e3fb80c11b
Type fulltextMimetype application/pdf
Order online >>

Other links

Publisher's full text

Search in DiVA

By author/editor
Klint, Elisabeth
By organisation
Division of Biomedical EngineeringFaculty of Science & Engineering
Radiology and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 609 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

doi
isbn
urn-nbn

Altmetric score

doi
isbn
urn-nbn
Total: 1056 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf