Open this publication in new window or tab >>2025 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]
This thesis explores ways of long-term monitoring of laboratory methods and their clinical applicability. N-terminal pro-B-type natriuretic peptide (NT-proBNP) and albumin are two analytes with general usage in the healthcare system where quality aspects play a major role in patient management. Data from laboratory information systems and external quality assessment schemes was utilised to explore analytical precision and accuracy between different methods and to what extent these parameters need to be considered when drawing different medical boundaries.
Study I showed that method bias has been increasing in a more heterogeneous market with manufacturers measuring NT-proBNP at different levels, which is problematic since international guidelines do not take method differences into account. Manufacturers generally meet the recommendation of imprecision <10%, although at NT-proBNP <125 ng/L imprecision reaches ≥10% more often.
Study II showed that central laboratory methods generally perform better than point-of-care instruments regarding method bias and imprecision. Standardisation efforts for urine creatinine have improved the situation, but further efforts are still needed for urine albumin, where certified reference material for serum has been available since 2008. Even though albumin methods consistently measure at different levels, recommendations on method bias and imprecision for classifying albuminuria are generally met.
Study III showed that patient median results at large could replicate the findings from study I based on external quality assessment data. Patient median results are a relevant and readily available tool that can add supplementary information about, for instance, commutability issues to internal and external controls. NT-proBNP requests have been steadily increasing, and usage of NT-proBNP differs between counties.
Study IV showed that NT-proBNP increases with age and is moderately negatively correlated with left ventricular ejection fraction (LVEF), but due to significant variability, it is unlikely that a single NT-proBNP value could accurately predict LVEF on an individual level. NT-proBNP performed better than LVEF in predicting the presence of a heart failure diagnosis.
These findings show that long-term monitoring of laboratory methods can be successfully achieved with, for instance, external controls and patient result medians, which also add information about clinical applicability, such as usage in accordance with guidelines.
Place, publisher, year, edition, pages
Uppsala: Acta Universitatis Upsaliensis, 2025. p. 76
Series
Digital Comprehensive Summaries of Uppsala Dissertations from the Faculty of Medicine, ISSN 1651-6206 ; 2189
Keywords
heart failure, immunochemistry, natriuretic peptides, quality control, reference standards
National Category
Clinical Laboratory Medicine
Research subject
Medical Science
Identifiers
urn:nbn:se:uu:diva-568160 (URN)978-91-513-2612-2 (ISBN)
Public defence
2025-11-24, Rosénsalen, Akademiska sjukhuset, Uppsala, 08:00 (Swedish)
Opponent
Supervisors
2025-10-312025-09-282025-10-31