Dimensionering av Södersjukhusets sterilcentral
KTH, School of Technology and Health (STH).
KTH, School of Technology and Health (STH).
2016 (Swedish)Independent thesis Basic level (degree of Bachelor), 10 credits / 15 HE creditsStudent thesisAlternative title
Dimensioning of the sterilization center at Södersjukhuset (English)
Abstract [sv]

I dagsläget pågår en omfattande planering av utbyggnationen samt upprustningen av Södersjukhuset. Byggnationen beräknas preliminärt vara i bruk under mars 2018. I och med förändringen kommer sjukhuset få en helt ny sterilcentral. 

Till en sterilcentral anländer kontaminerat medicinskt gods från tillhörande eller externa sjukhus. Efter att ha genomgått diskdesinficering och autoklavering på sterilcentralen kommer godset så småningom ut som sterilt för ytterligare användning. 

Det finns idag en i stort sett färdig ritning för den nya sterilcentralen, men en osäkerhet i om antal utritade diskdesinfektorer och autoklaver kommer kunna omhänderta det framtida flödet av kontaminerat medicinskt gods. Målet med projektet var att undersöka Södersjukhusets preliminära ritning och sedan avgöra om ritningens diskdesinfektor- samt autoklavlinje är rätt dimensionerad och kan följas vid utbyggnationen. För att uppfylla detta mål har flödesundersökningar samt kapacitetsberäkningar utförts på dagens sterilcentral. Resultaten från dessa undersökningar har sedan jämförts med ritningen för att kunna dra en slutsats i huruvida ritningens specifikationer gällande diskdesinfektor- samt autoklavlinje är tillräckliga. 

Antalet maskiner i den preliminära ritningen kommer kunna hantera dagens flöde av medicinskt gods om de har en kapacitet minst lika stor som dagens utrustning. Därför drogs slutsatsen att denna ritning kunde anses bli den slutgiltiga.

Abstract [en]

At the moment a comprehensive planning of the reconstruction of Södersjukhuset is in progress. The reconstruction is estimated to be done during March 2018, and with this the hospital will receive a new sterilization center.

A sterilization center takes care of medical material that is contaminated at arrival. After the sterilization process, including desinfection and autoclaving, the material leaves the sterilization center as sterile.

Today, a drawing of the new sterilization center exists, but there is an uncertainty whether the number of drawn washing desinfectors and autoclaves will be able to handle the incoming material. The goal with the project was to investigate the current drawing and then conclude whether the number of washing desinfectors and autoclaves is correctly dimensioned, and can be followed at the reconstruction. To reach the goal, investigations of the flow and calculations of the capacity, has been done at the current sterilization center. These have then been compared to the drawing, which made it possible to conclude whether the drawing’s number of machines is correctly dimensioned.

The number of machines in the present drawing will be able to take care of the daily flow of medical material, if they have a capacity of at least the same as the current machines. Thus the conclusion came to be that the drawing can be considered as the definitive.

Place, publisher, year, edition, pages
2016. , 28 p.
Series
TRITA-STH, 2016:29
Keyword [sv]
Sterilcentraler, Logistik, Kemiska processer, Sterilteknik, Flöde
National Category
Medical Engineering
Identifiers
URN: urn:nbn:se:kth:diva-190733OAI: oai:DiVA.org:kth-190733DiVA: diva2:952496
Subject / course
Medical Engineering
Educational program
Master of Science in Engineering - Medical Engineering
Supervisors
Examiners
Available from: 2016-08-15 Created: 2016-08-14 Last updated: 2016-08-15Bibliographically approved

Open Access in DiVA

fulltext(1728 kB)13 downloads
File information
File name FULLTEXT01.pdfFile size 1728 kBChecksum SHA-512
7c622c943d01118037b7eecddbf15ce087e2bb89532e82681e0b58a25d9c18faf94740b5d71b0fce9c588e4e5bb3ef124f2fc7671e4f6c6b1dc1746706697be9
Type fulltextMimetype application/pdf

By organisation
School of Technology and Health (STH)
Medical Engineering

Search outside of DiVA

GoogleGoogle Scholar
Total: 13 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available