On the use of computed tomography in cardiac imaging
Linköping University, Department of Medical and Health Sciences, Division of Radiological Sciences. Linköping University, Faculty of Medicine and Health Sciences. Linköping University, Center for Medical Image Science and Visualization (CMIV).
2016 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Background

Cardiac Computed Tomography Angiography (CCTA) is becoming increasingly useful in the work‐up of coronary artery disease (CAD). Several potential methods for increasing the diagnostic yield of cardiac CT are available.

Purpose

Study I: To investigate whether the use of a 2‐D, non‐linear adaptive noise reduction filter can improve CCTA image quality.

Study II: To evaluate the variation in adenosine stress dynamic CT perfusion (CTP) blood flow as compared to stress 99mTc SPECT. Secondly, to compare the perfusion results from manual and automatic myocardial CTP segmentation.

Study III: To evaluate the accuracy of non‐invasive, CCTA‐derived Fractional Flow Reserve (cFFR).

Study IV: To evaluate the prognostic value of CCTA in terms of major adverse cardiac events (MACE).

Materials and methods

Study I: Single images from 36 consecutive CCTA exams performed with two different dose levels were used. Image quality in full dose, low‐dose and noise‐reduced low‐dose images was graded using visual grading analysis. Image noise was measured.

Study II: CTP and SPECT were performed in 17 patients, and the variation in per AHA‐segment blood flow was evaluated and compared. CTP results from manual and automated image segmentation were compared.

Study III: CCTA datasets from 21 patients were processed using cFFR software and the results compared to the corresponding invasively measured FFR (invFFR).

Study IV: 1205 consecutive patients with chest pain of unknown origin underwent CCTA. Baseline data and data on subsequent MACE were retrieved from relevant registries. Survival, hazard ratios and the three‐year incidence of cardiac events and readmissions were calculated.

Results

Study I: There was significant improvement in perceived image quality for all criteria when the filter was applied, and a significant decrease in image noise.

Study II: The correlation coefficients for CTP vs. SPECT were 0.38 and 0.41 (p<0.001, for manual and automated segmentation respectively. Mean per patient CTP blood flow in normal segments varied between 94‐183 ml/100 ml tissue/min for manual segmentation, and 104‐196 ml/100 ml tissue/min for automated segmentation. The Spearman rank correlation coefficient for manual vs. automated segmentation CTP was ρ = 0.88 (p<0.001) and the Intraclass Correlation Coefficient (ICC) was 0.93 (p<0.001).

Study III: The Spearman rank correlation coefficient for cFFR vs. invFFR was ρ = 0.77 (p<0.001) and the ICC was 0.73 (p<0.001). Sensitivity, specificity, positive predictive value and negative predictive value for significant stenosis (FFR<0.80, per vessel) were 0.83, 0.76, 0.56 and 0.93 respectively.

Study IV: The hazard ratio for non‐obstructive CAD vs. normal coronary arteries was 5.13 (95% C.I 1.03‐25.43, p<0.05), and 151.40 (95% C.I 37.03‐619.08, p<0.001) for obstructive CAD vs. normal coronary arteries. The three‐year incidence of MACE was 1.1% for patients with normal vessels on CCTA, 2.5% for patients with non‐obstructive CAD and 42.7% for patients with obstructive CAD (p<0.001).

Conclusions:

Study I: Image quality and noise levels of low dose images were significantly improved with the filter, even though the improvement was small compared to the image quality of the corresponding diastolic full‐dose images.

Study II: Correlation between dynamic CTP and SPECT was positive but weak. There were large variations in CTP blood flow in normal segments on SPECT, rendering the definition of an absolute cut‐off value for normal vs. ischemic myocardium difficult. Manual and automatic segmentation were equally useful.

Study III: The correlation between cFFR and invFFR was good, indicating that noninvasively estimated cFFR performs on a similar level as invasively measure FFR.

Study IV: The long‐term risk for MACE was very low in patients without obstructive CAD on CCTA, though there seemed to be a substantial increase in the risk for MACE even in patients with non‐obstructive CAD as compared to normal coronary arteries. In addition, even patients with normal coronary arteries or non‐obstructive CAD continued to have a substantial number of readmissions for chest pain or angina pectoris.

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press, 2016. , 73 p.
Series
Linköping University Medical Dissertations, ISSN 0345-0082 ; 1518
National Category
Radiology, Nuclear Medicine and Medical Imaging Medical Image Processing Cardiac and Cardiovascular Systems
Identifiers
URN: urn:nbn:se:liu:diva-128276DOI: 10.3384/diss.diva-128276ISBN: 978-91-7685-795-3 (print)OAI: oai:DiVA.org:liu-128276DiVA: diva2:930557
Public defence
2016-06-09, Berzeliussalen, ingång 65, plan, 9, Campus US, Linköping, 09:00 (Swedish)
Opponent
Supervisors
Funder
Region ÖstergötlandSwedish Heart Lung Foundation, 20120449
Available from: 2016-05-24 Created: 2016-05-24 Last updated: 2016-05-25Bibliographically approved
List of papers
1. The efficacy of 2D, non-linear noise reduction filtering in cardiac imaging: a pilot study
Open this publication in new window or tab >>The efficacy of 2D, non-linear noise reduction filtering in cardiac imaging: a pilot study
2011 (English)In: Acta Radiologica, ISSN 0284-1851, E-ISSN 1600-0455, Vol. 52, no 7, 716-722 p.Article in journal (Refereed) Published
Abstract [en]

Background: Computed tomography (CT) is becoming increasingly popular as a non-invasive method for visualizing the coronary arteries but patient radiation doses are still an issue. Postprocessing filters such as 2D adaptive non-linear filters might help to reduce the dose without loss of image quality. less thanbrgreater than less thanbrgreater thanPurpose: To investigate whether the use of a 2D, non-linear adaptive noise reduction filter can improve image quality in cardiac computed tomography angiography (CCTA). less thanbrgreater than less thanbrgreater thanMaterial and Methods: CCTA examinations were performed in 36 clinical patients on a dual source CT using two patient dose levels: maximum dose during diastole and reduced dose (20% of maximum dose) during systole. One full-dose and one reduced-dose image were selected from each of the examinations. The reduced-dose image was duplicated and one copy postprocessed using a 2D non-linear adaptive noise reduction filter, resulting in three images per patient. Image quality was assessed using visual grading with three criteria from the European guidelines for assessment of image quality and two additional criteria regarding the left main artery and the overall image quality. Also, the HU value and its standard deviation were measured in the ascending and descending aorta. Data were analyzed using Visual Grading Regression and paired t-test. less thanbrgreater than less thanbrgreater thanResult: For all five criteria, there was a significant (P andlt; 0.01 or better) improvement in perceived image quality when comparing postprocessed low-dose images with low-dose images without noise reduction. Comparing full dose images with postprocessed low-dose images resulted in a considerably larger, significant (P andlt; 0.001) difference. Also, there was a significant reduction of the standard deviation of the HU values in the ascending and descending aorta when comparing postprocessed low-dose images with low-dose images without postprocessing. less thanbrgreater than less thanbrgreater thanConclusion: Even with an 80% dose reduction, there was a significant improvement in the perceived image quality when using a 2D noise-reduction filter, though not approaching the quality of full-dose images. This indicates that cardiac CT examinations could benefit from noise-reducing postprocessing with 2D non-linear adaptive filters.

Place, publisher, year, edition, pages
Informa Healthcare / Wiley-Blackwell / Royal Society of Medicine Press, 2011
Keyword
Cardiac, CT angiography, heart, adults, image manipulation
National Category
Medical and Health Sciences
Identifiers
urn:nbn:se:liu:diva-71803 (URN)10.1258/ar.2011.100511 (DOI)000295759600007 ()
Available from: 2011-11-04 Created: 2011-11-04 Last updated: 2017-12-08
2. Large variation in blood flow between left ventricular segments, as detected by adenosine stress dynamic CT perfusion.
Open this publication in new window or tab >>Large variation in blood flow between left ventricular segments, as detected by adenosine stress dynamic CT perfusion.
Show others...
2015 (English)In: Clinical Physiology and Functional Imaging, ISSN 1475-0961, E-ISSN 1475-097X, Vol. 35, no 4, 291-300 p.Article in journal (Refereed) Published
Abstract [en]

BACKGROUND: Dynamic cardiac CT perfusion (CTP) is based on repeated imaging during the first-pass contrast agent inflow. It is a relatively new method that still needs validation.

PURPOSE: To evaluate the variation in adenosine stress dynamic CTP blood flow as compared to (99m) Tc SPECT. Secondarily, to compare manual and automatic segmentation.

METHODS: Seventeen patients with manifest coronary artery disease were included. Nine were excluded from evaluation for various reasons. All patients were examined with dynamic stress CTP and stress/rest SPECT. CTP blood flow was compared with SPECT on a per segment basis. Results for manual and automated AHA segmentation were compared.

RESULTS: CTP showed a positive correlation with SPECT, with correlation coefficients of 0·38 and 0·41 for manual and automatic segmentation, respectively (P<0·0001). There was no significant difference between the correlation coefficients of the manual and automated segmentation procedures (P = 0·75). The average per individual global CTP blood flow value for normal segments varied by a factor of 1·9 (manual and automatic segmentation). For the whole patient group, the CTP blood flow value in normal segments varied by a factor of 2·9/2·7 (manual/automatic segmentation). Within each patient, the average per segment blood flow in normal segments varied by a factor of 1·3-2·0/1·2-2·1 (manual/automatic segmentation).

CONCLUSION: A positive but rather weak correlation was found between CTP and (99m) Tc SPECT. Large variations in CTP blood flow suggest that a cut-off value for stress myocardial blood flow is inadequate to detect ischaemic segments. Dynamic CTP is hampered by a limited coverage.

National Category
Clinical Medicine
Identifiers
urn:nbn:se:liu:diva-113400 (URN)10.1111/cpf.12163 (DOI)000356312800007 ()24842265 (PubMedID)
Available from: 2015-01-17 Created: 2015-01-17 Last updated: 2017-12-05
3. Software-based on-site estimation of fractional flow reserve using standard coronary CT angiography data.
Open this publication in new window or tab >>Software-based on-site estimation of fractional flow reserve using standard coronary CT angiography data.
Show others...
2016 (English)In: Acta Radiologica, ISSN 0284-1851, E-ISSN 1600-0455, Vol. 57, no 10, 1186-1192 p.Article in journal (Refereed) Published
Abstract [en]

BACKGROUND: The significance of a coronary stenosis can be determined by measuring the fractional flow reserve (FFR) during invasive coronary angiography. Recently, methods have been developed which claim to be able to estimate FFR using image data from standard coronary computed tomography angiography (CCTA) exams.

PURPOSE: To evaluate the accuracy of non-invasively computed fractional flow reserve (cFFR) from CCTA.

MATERIAL AND METHODS: A total of 23 vessels in 21 patients who had undergone both CCTA and invasive angiography with FFR measurement were evaluated using a cFFR software prototype. The cFFR results were compared to the invasively obtained FFR values. Correlation was calculated using Spearman's rank correlation, and agreement using intraclass correlation coefficient (ICC). Sensitivity, specificity, accuracy, negative predictive value, and positive predictive value for significant stenosis (defined as both FFR ≤0.80 and FFR ≤0.75) were calculated.

RESULTS: The mean cFFR value for the whole group was 0.81 and the corresponding mean invFFR value was 0.84. The cFFR sensitivity for significant stenosis (FFR ≤0.80/0.75) on a per-lesion basis was 0.83/0.80, specificity was 0.76/0.89, and accuracy 0.78/0.87. The positive predictive value was 0.56/0.67 and the negative predictive value was 0.93/0.94. The Spearman rank correlation coefficient was ρ = 0.77 (P < 0.001) and ICC = 0.73 (P < 0.001).

CONCLUSION: This particular CCTA-based cFFR software prototype allows for a rapid, non-invasive on-site evaluation of cFFR. The results are encouraging and cFFR may in the future be of help in the triage to invasive coronary angiography.

Place, publisher, year, edition, pages
Sage Publications, 2016
Keyword
Cardiac; computed tomography angiography (CTA); heart; arteries; adults; computer applications – detection/diagnosis
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
urn:nbn:se:liu:diva-123579 (URN)10.1177/0284185115622075 (DOI)000382967500007 ()26691914 (PubMedID)
Note

Funding agencies: Department of Radiology, Region Ostergotland; Swedish Heart-Lung-foundation [20120449]

Available from: 2015-12-29 Created: 2015-12-29 Last updated: 2017-12-01Bibliographically approved

Open Access in DiVA

fulltext(2780 kB)129 downloads
File information
File name FULLTEXT01.pdfFile size 2780 kBChecksum SHA-512
f135286f6f737a7810272d5d8291638895fb364fb3291ca76631aa10cf3735bf0dff3d4789bef1fb38fcdc6a30e564f786a3af1ca4f8683b89b633c95dac8ef2
Type fulltextMimetype application/pdf
omslag(434 kB)20 downloads
File information
File name COVER01.pdfFile size 434 kBChecksum SHA-512
fc51a3d577f72585c25d2a55071eadc26c5cbe2e59e84a9c44bca256b7d44647196dba072fdaa995de4dfdd75707a32d796b625dfe8242edc15c07e85b6ae532
Type coverMimetype application/pdf

Other links

Publisher's full text

Search in DiVA

By author/editor
De Geer, Jakob
By organisation
Division of Radiological SciencesFaculty of Medicine and Health SciencesCenter for Medical Image Science and Visualization (CMIV)
Radiology, Nuclear Medicine and Medical ImagingMedical Image ProcessingCardiac and Cardiovascular Systems

Search outside of DiVA

GoogleGoogle Scholar
Total: 129 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

doi
isbn
urn-nbn

Altmetric score

doi
isbn
urn-nbn
Total: 1376 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf