Size-dependent scanning parameters (kVp and mAs) for photon-counting spectral CT system in pediatric imaging: simulation study
KTH, School of Engineering Sciences (SCI), Physics, Physics of Medical Imaging. (Physics of Medical imaging)ORCID iD: 0000-0001-8560-3262
KTH, School of Engineering Sciences (SCI), Physics, Physics of Medical Imaging.
KTH, School of Engineering Sciences (SCI), Physics, Physics of Medical Imaging.
2016 (English)In: Physics in Medicine and Biology, ISSN 0031-9155, E-ISSN 1361-6560, Vol. 61, no 11Article in journal (Refereed) Published
Abstract [en]

We are developing a photon-counting spectral CT detector with small pixel size of 0.40.5 mm2, o ering a potentialadvantage for better visualization of small structures in pediatric patients. The purpose of this study is to determinethe patient size dependent scanning parameters (kVp and mAs) for pediatric CT in two imaging cases: adipose imagingand iodinated blood imaging.Cylindrical soft-tissue phantoms of diameters between 10-25 cm were used to mimic patients of di erent ages from 0-15 y. For adipose imaging, a 5-mm-diameter adipose sphere was assumed as an imaging target, while an iodinated bloodsphere of 1 mm in diameter was assumed in the case of iodinated imaging. By applying the geometry of a commercial CTscanner (GE LightSpeed VCT), simulations were carried out to calculate the detectability index,d02, with tube potentialsvarying from 40 to 140 kVp. The optimal kVp for each phantom in each imaging case was determined such that the dose-normalized detectability index,d02=dose, is maximized. With the assumption that image quality in pediatric imagingis required the same as in typical adult imaging, the value of mAs at optimal kVp for each phantom was selected toachieve a reference detectability index that was obtained by scanning an adult phantom (30 cm in diameter) in a typicaladult CT procedure (120 kVp and 200 mAs) using a modeled energy-integrating system.For adipose imaging, the optimal kVps are 50, 60, 80, and 120 kVp, respectively, for phantoms of 10, 15, 20, and25-cm in diameter. The corresponding mAs values required to achieve the reference detectability index are only 9%,23%, 24%, and 54% of the mAs that is used for adult patients at 120 kVp, for 10, 15, 20, and 25-cm-diameter phantoms,respectively. In the case of iodinated imaging, a tube potential of 60 kVp was found optimal for all phantoms investigated,and the mAs values required to achieve the reference detectability index are 2%, 9%, 37%, and 109% of the adult mAs.The results also indicate that with the use of respective optimal kVps, the photon-counting spectral system o ers up to30% higherd02=dose than the modeled energy-integrating system for adipose imaging, and 70% for iodinated imaging.

Place, publisher, year, edition, pages
Institute of Physics Publishing (IOPP), 2016. Vol. 61, no 11
Keyword [en]
photon counting, spectral CT, silicon-strip detector, pediatric imaging, detectability index
National Category
Radiology, Nuclear Medicine and Medical Imaging
URN: urn:nbn:se:kth:diva-184605DOI: 10.1088/0031-9155/61/11/4105ISI: 000377427700014ScopusID: 2-s2.0-84971515147OAI: diva2:916328

QC 20160706

Available from: 2016-04-01 Created: 2016-04-01 Last updated: 2016-07-06Bibliographically approved
In thesis
1. Characterization and Optimization of Silicon-strip Detectors for Mammography and Computed Tomography
Open this publication in new window or tab >>Characterization and Optimization of Silicon-strip Detectors for Mammography and Computed Tomography
2016 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

The goal in medical x-ray imaging is to obtain the image quality requiredfor a given detection task, while ensuring that the patient dose is kept as lowas reasonably achievable. The two most common strategies for dose reductionare: optimizing incident x-ray beams and utilizing energy informationof transmitted beams with new detector techniques (spectral imaging). Inthis thesis, dose optimization schemes were investigated in two x-ray imagingsystems: digital mammography and computed tomography (CT).

In digital mammography, the usefulness of anti-scatter grids was investigatedas a function of breast thickness with varying geometries and experimentalconditions. The general conclusion is that keeping the grid is optimalfor breasts thicker than 5 cm, whereas the dose can be reduced without a gridfor thinner breasts.

A photon-counting silicon-strip detector developed for spectral mammographywas characterized using synchrotron radiation. Energy resolution, ΔE/Ein, was measured to vary between 0.11-0.23 in the energy range 15-40 keV, which is better than the energy resolution of 0.12-0.35 measured inthe state-of-the-art photon-counting mammography system. Pulse pileup hasshown little effect on energy resolution.

In CT, the performance of a segmented silicon-strip detector developedfor spectral CT was evaluated and a theoretical comparison was made withthe state-of-the-art CT detector for some clinically relevant imaging tasks.The results indicate that the proposed photon-counting silicon CT detector issuperior to the state-of-the-art CT detector, especially for high-contrast andhigh-resolution imaging tasks.

The beam quality was optimized for the proposed photon-counting spectralCT detector in two head imaging cases: non-enhanced imaging and Kedgeimaging. For non-enhanced imaging, a 120-kVp spectrum filtered by 2half value layer (HVL) copper (Z = 29) provides the best performance. Wheniodine is used in K-edge imaging, the optimal filter is 2 HVL iodine (Z = 53)and the optimal kVps are 60-75 kVp. In the case of gadolinium imaging, theradiation dose can be minimized at 120 kVp filtered by 2 HVL thulium (Z =69).

Place, publisher, year, edition, pages
STOCKHOLM: KTH Royal Institute of Technology, 2016. viii, 73 p.
TRITA-FYS, ISSN 0280-316X ; 0280-316X
mammography, anti-scatter grid, photon-counting, spectral computed tomography, silicon strip, ASIC, energy resolution, Compton scatter, material decomposition, K-edge imaging
National Category
Medical Engineering
Research subject
Medical Technology
urn:nbn:se:kth:diva-184092 (URN)978-91-7595-919-1 (ISBN)
Public defence
2016-04-22, FA 31, ROSLAGSTULLSBACKEN 21, KTH, STOCKHOLM, 09:00 (English)

QC 20160401

Available from: 2016-04-01 Created: 2016-03-23 Last updated: 2016-04-01Bibliographically approved

Open Access in DiVA

fulltext(718 kB)47 downloads
File information
File name FULLTEXT01.pdfFile size 718 kBChecksum SHA-512
Type fulltextMimetype application/pdf

Other links

Publisher's full textScopus

Search in DiVA

By author/editor
Chen, HanDanielsson, MatsXu, Cheng
By organisation
Physics of Medical Imaging
In the same journal
Physics in Medicine and Biology
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 47 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 85 hits
ReferencesLink to record
Permanent link

Direct link