Change search
ReferencesLink to record
Permanent link

Direct link
Properties of screen printed electrocardiography smartware electrodes investigated in an electro-chemical cell
Linköping University, Department of Biomedical Engineering, Physiological Measurements. Linköping University, The Institute of Technology.
Department of Materials Chemistry, Uppsala, Sweden.
Acreo AB, Norrköping, Sweden.
Acreo AB, Norrköping, Sweden.
Show others and affiliations
2013 (English)In: Biomedical engineering online, ISSN 1475-925X, Vol. 12Article in journal (Refereed) Published
Abstract [en]


ECG (Electrocardiogram) measurements in home health care demands new sensor solutions. In this study, six different configurations of screen printed conductive ink electrodes have been evaluated with respect to electrode potential variations and electrode impedance.


The electrode surfaces consisted of a Ag/AgCl-based ink with a conduction line of carbon or Ag-based ink underneath. On top, a lacquer layer was used to define the electrode area and to cover the conduction lines. Measurements were performed under well-defined electro-chemical conditions in a physiologic saline solution.


The results showed that all printed electrodes were stable and have a very small potential drift (less than 3 mV/30 min). The contribution to the total impedance was 2% of the set maximal allowed impedance (maximally 1 kΩ at 50 Hz), assuming common values of input impedance and common mode rejection ratio of a regular amplifier.


Our conclusions are that the tested electrodes show satisfying properties to be used as elements in a skin electrode design that could be suitable for further investigations by applying the electrodes on the skin.

Place, publisher, year, edition, pages
2013. Vol. 12
Keyword [en]
Screen printed electrodes, ECG, Electrode impedance, Electrode potential, Smartware electrodes
National Category
Engineering and Technology
URN: urn:nbn:se:liu:diva-96422DOI: 10.1186/1475-925X-12-64ISI: 000321916300001OAI: diva2:641946

Funding Agencies|VINNOVA - Swedens Innovation Agency||NovaMedTech||Linkoping Initiative for Life Science Technologies (LIST)||

Available from: 2013-08-20 Created: 2013-08-19 Last updated: 2013-10-30
In thesis
1. Smartware electrodes for ECG measurements: Design, evaluation and signal processing
Open this publication in new window or tab >>Smartware electrodes for ECG measurements: Design, evaluation and signal processing
2013 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

The aim of this thesis work has been to study textile and screen printed smartware electrodes for electrocardiographic (ECG) measurements both in terms of their electrode properties and possibility to further improve their robustness to movement induced noise by using signal processing. Smartware electrodes for ECG measurements have previously been used in various applications but basic electrical electrode properties have not sufficiently been looked into. Furthermore, we believe that there is a possibility to reduce disturbances in the smartware ECG by adding redundant sensors and applying sensor fusion signal processing.

Electrical properties of conductive textiles have been evaluated in terms of stability and electrode impedance. Three yarns and textile electrode surfaces were tested. The electrodes made from pure stainless steel and 50\% stainless steel/ 50\% polyester showed acceptable stability of electrode potentials. All electrode measurements were performed on skin.

Furthermore, we produced six screen printed electrodes and their electrical performance was investigated in an electrochemical cell. The tested inks contained carbon or silver particles in the conduction lines, and Ag/AgCl particles in the electrode surface. Results show that all electrodes were stable in time, with a maximum drift of a few mV during 30 minutes. The silver ink is superior to the carbon based in terms of electrode impedance at the higher frequencies.

To extract viable physiological information from noisy signals, canonical correlation analysis (CCA) was applied on multi-channel ECG signals recorded with textile electrodes. Using CCA to solve the blind source separation (BSS) problem, we intended to separate the ECG signal from the various noise sources. The method (CCABSS) was compared to averaging of the ECG channels and to the independent component analysis method (ICA). In the dataset consisting of noisy ECG recordings, the signal was uninterpretable in 7% after CCABSS. Corresponding values for averaging and ICA were 33% and 17%, respectively.

Smartware applications often include heartbeat detection while moving, a measurement situation which is prone to produce noise corrupted ECG signals. To compensate for this, we used an event detector based on a multi-channel input, a model of the event and weighted correlation. For measurements at rest and static muscle tension, the sensitivity of the event detector was 97% and 77% respectively. Corresponding values for the golden standard detector Pan-Tompkins were 96% and 52%, respectively.


Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press, 2013. 91 p.
Linköping Studies in Science and Technology. Dissertations, ISSN 0345-7524 ; 1546
National Category
Other Medical Engineering
urn:nbn:se:liu:diva-100134 (URN)10.3384/diss.diva-100134 (DOI)978-91-7519-507-0 (print) (ISBN)
Public defence
2013-12-13, Eken, Campus US, Linköpings universitet, Linköping, 10:00 (Swedish)
Available from: 2013-10-30 Created: 2013-10-29 Last updated: 2013-11-04Bibliographically approved

Open Access in DiVA

fulltext(476 kB)228 downloads
File information
File name FULLTEXT01.pdfFile size 476 kBChecksum SHA-512
Type fulltextMimetype application/pdf

Other links

Publisher's full text

Search in DiVA

By author/editor
Rattfält, LindaAsk, Per
By organisation
Physiological MeasurementsThe Institute of Technology
In the same journal
Biomedical engineering online
Engineering and Technology

Search outside of DiVA

GoogleGoogle Scholar
Total: 228 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 383 hits
ReferencesLink to record
Permanent link