Improved calibration procedure for laser Doppler perfusion monitors
Linköping University, Department of Biomedical Engineering, Biomedical Instrumentation. Linköping University, The Institute of Technology. (Perimed AB, Järfälla-Stockholm, Sweden)ORCID iD: 0000-0002-3454-6576
Linköping University, Department of Biomedical Engineering, Biomedical Instrumentation. Linköping University, The Institute of Technology. Linköping University, Faculty of Health Sciences.ORCID iD: 0000-0001-6385-6760
Perimed AB, Järfälla-Stockholm, Sweden.
Linköping University, Department of Biomedical Engineering, Biomedical Instrumentation. Linköping University, The Institute of Technology.
2011 (English)In: Optical Diagnostics and SensingXI: Toward Point-of-CareDiagnostics; and Design andPerformance Validation ofPhantoms Used in Conjunctionwith Optical Measurement ofTissue III / [ed] Robert J. Nordstrom; Gerard L. Coté, SPIE - International Society for Optical Engineering, 2011, 790602-1-790602-7 p.Conference paper (Other academic)
Abstract [en]

Commercial laser Doppler perfusion monitors are calibrated using the perfusion value, i.e. the first order moment of the Doppler power spectrum, from a measurement in a standardized microsphere colloidal suspension under Brownian motion. The calibration perfusion value depends on several parameters of the suspension that are difficult to keep constant with adequate accuracy, such as the concentration, temperature and the microsphere size distribution. The calibration procedure itself may therefore introduce significant errors in the measured values.

An altered calibration procedure, where the zero order moment is used is described and demonstrated in this paper. Since the above mentioned parameters only affect the frequency content of the Doppler power spectrum and not the total power, the zero order moment will be independent of those parameters. It is shown that the variation in the calibration value, as given by measurements on different scattering liquids with a wide range of scattering properties and temperatures, is only a few percent using the proposed method. For the conventional calibration procedure, this variation corresponds to an error introduced by merely a 1°C variation in the reference liquid temperature. The proposed calibration method also enables absolute level comparisons between measured and simulated Doppler power spectra.

Place, publisher, year, edition, pages
SPIE - International Society for Optical Engineering, 2011. 790602-1-790602-7 p.
Series
, Proceedings of SPIE (Progress in biomedical optics and imaging), ISSN 1605-7422 ; Vol. 7906
Keyword [en]
Laser Doppler flowmetry, calibration, Brownian motion, optical phantoms
National Category
Medical Laboratory and Measurements Technologies
Identifiers
URN: urn:nbn:se:liu:diva-66091DOI: 10.1117/12.871938ISI: 000302694400002ISBN: 9780819484437OAI: oai:DiVA.org:liu-66091DiVA: diva2:401575
Conference
Optical Diagnostics and Sensing XI: Toward Point-of-Care Diagnostics; and Design and Performance Validation of Phantoms Used in Conjunction with Optical Measurement of Tissue III, 22-26 January 2011, 22-24 January 2011, San Francisco, CA, USA
Note

Original Publication: Ingemar Fredriksson, M. Larsson, T. Strömberg and F. Salomonsson, Improved calibration procedure for laser Doppler perfusion monitors, 2011. http://dx.doi.org/10.1117/12.871938 Copyright 2011 Society of Photo-Optical Instrumentation Engineers. One print or electronic copy may be made for personal use only. Systematic reproduction and distribution, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited.

The publication is also included in the series Progress in Biomedical Optics and Imaging (ISSN 1605-7422) vol. 12 issue 24.

Available from: 2011-03-07 Created: 2011-03-03 Last updated: 2016-08-31Bibliographically approved

Open Access in DiVA

Improved calibration procedure for laser Doppler perfusion monitors(388 kB)705 downloads
File information
File name FULLTEXT02.pdfFile size 388 kBChecksum SHA-512
e5f4ef614ca94cff0553237af63694c7e1bcadb802fb00ec5903529a7297da16ee4894c41334185db0619ebbf5e7be7196f34b63f7411b95c560b22587247d74
Type fulltextMimetype application/pdf

Other links

Publisher's full text

Search in DiVA

By author/editor
Fredriksson, IngemarLarsson, MarcusStrömberg, Tomas
By organisation
Biomedical InstrumentationThe Institute of TechnologyFaculty of Health Sciences
Medical Laboratory and Measurements Technologies

Search outside of DiVA

GoogleGoogle Scholar
Total: 705 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 443 hits
ReferencesLink to record
Permanent link

Direct link