Testing the methodology for a dosimetric end-to-end audit of IMRT/VMAT: results of IAEA multicentre and national studies
IAEA, Austria.
Med Univ Vienna, Austria; Christian Doppler Lab Med Radiat Res Radiat Oncol, Austria.
Med Univ Vienna, Austria; Christian Doppler Lab Med Radiat Res Radiat Oncol, Austria.
IAEA, Austria.
Show others and affiliations
2019 (English)In: Acta Oncologica, ISSN 0284-186X, E-ISSN 1651-226XArticle in journal (Refereed) Epub ahead of print
Abstract [en]

Introduction: Within an International Atomic Energy Agency (IAEA) co-ordinated research project (CRP), a remote end-to-end dosimetric quality audit for intensity modulated radiation therapy (IMRT)/ volumetric arc therapy (VMAT) was developed to verify the radiotherapy chain including imaging, treatment planning and dose delivery. The methodology as well as the results obtained in a multicentre pilot study and national trial runs conducted in close cooperation with dosimetry audit networks (DANs) of IAEA Member States are presented. Material and methods: A solid polystyrene phantom containing a dosimetry insert with an irregular solid water planning target volume (PTV) and organ at risk (OAR) was designed for this audit. The insert can be preloaded with radiochromic film and four thermoluminescent dosimeters (TLDs). For the audit, radiotherapy centres were asked to scan the phantom, contour the structures, create an IMRT/VMAT treatment plan and irradiate the phantom. The dose prescription was to deliver 4 Gy to the PTV in two fractions and to limit the OAR dose to a maximum of 2.8 Gy. The TLD measured doses and film measured dose distributions were compared with the TPS calculations. Results: Sixteen hospitals from 13 countries and 64 hospitals from 6 countries participated in the multicenter pilot study and in the national runs, respectively. The TLD results for the PTV were all within +/- 5% acceptance limit for the multicentre pilot study, whereas for national runs, 17 participants failed to meet this criterion. All measured doses in the OAR were below the treatment planning constraint. The film analysis identified seven plans in national runs below the 90% passing rate gamma criteria. Conclusion: The results proved that the methodology of the IMRT/VMAT dosimetric end-to-end audit was feasible for its intended purpose, i.e., the phantom design and materials were suitable; the phantom was easy to use and it was robust enough for shipment. Most importantly the audit methodology was capable of identifying suboptimal IMRT/VMAT delivery.

Place, publisher, year, edition, pages
TAYLOR & FRANCIS LTD , 2019.
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
URN: urn:nbn:se:liu:diva-160165DOI: 10.1080/0284186X.2019.1648859ISI: 000482281100001PubMedID: 31423867OAI: oai:DiVA.org:liu-160165DiVA, id: diva2:1349562
Available from: 2019-09-09 Created: 2019-09-09 Last updated: 2019-11-27

Open Access in DiVA

fulltext(2304 kB)5 downloads
File information
File name FULLTEXT01.pdfFile size 2304 kBChecksum SHA-512
e7fb8db6994c3b6f26a675b86aeaba370fedd94360a839c8f749e461605a20eeccc3a6bce409f311f34bb4ae2f9e36db558d977bc979b2da3a383d47d81c6665
Type fulltextMimetype application/pdf

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Carlsson Tedgren, ÅsaAdolfsson, Emelie
By organisation
Division of Radiological SciencesFaculty of Medicine and Health SciencesDepartment of Radiation Physics
In the same journal
Acta Oncologica
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 5 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 9 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf