ICL-based TDLAS sensor for real-time breath gas analysis of carbon monoxide isotopes
Umeå University, Faculty of Science and Technology, Department of Applied Physics and Electronics. (Applied Laser Spectroscopy)
Umeå University, Faculty of Science and Technology, Department of Applied Physics and Electronics. (Applied Laser Spectroscopy, TEC-Lab)ORCID iD: 0000-0002-5065-7786
2017 (English)In: Optics Express, ISSN 1094-4087, E-ISSN 1094-4087, Vol. 25, no 11, p. 12743-12752Article in journal (Refereed) Published
Abstract [en]

We present a compact sensor for carbon monoxide (CO) in air and exhaled breath based on a room temperature interband cascade laser (ICL) operating at 4.69 µm, a low-volume circular multipass cell and wavelength modulation absorption spectroscopy. A fringe-limited (1σ) sensitivity of 6.5 × 10−8 cm−1Hz-1/2 and a detection limit of 9 ± 5 ppbv at 0.07 s acquisition time are achieved, which constitutes a 25-fold improvement compared to direct absorption spectroscopy. Integration over 10 s increases the precision to 0.6 ppbv. The setup also allows measuring the stable isotope 13CO in breath. We demonstrate quantification of indoor air CO and real-time detection of CO expirograms from healthy non-smokers and a healthy smoker before and after smoking. Isotope ratio analysis indicates depletion of 13CO in breath compared to natural abundance.

Place, publisher, year, edition, pages
Optical Society of America, 2017. Vol. 25, no 11, p. 12743-12752
Keywords [en]
Laser sensors, Spectroscopy - infrared, Spectroscopy - laser, Biological sensing and sensors, Semiconductor lasers - quantum cascade
National Category
Medical Laboratory and Measurements Technologies Atom and Molecular Physics and Optics Physical Chemistry Physiology
Identifiers
URN: urn:nbn:se:umu:diva-135377DOI: 10.1364/OE.25.012743ISI: 000403940700064OAI: oai:DiVA.org:umu-135377DiVA, id: diva2:1098658
Available from: 2017-05-24 Created: 2017-05-24 Last updated: 2018-09-26Bibliographically approved
In thesis
1. Real-time breath gas analysis of carbon monoxide: laser-based detection and pulmonary gas exchange modeling
Open this publication in new window or tab >>Real-time breath gas analysis of carbon monoxide: laser-based detection and pulmonary gas exchange modeling
2018 (English)Doctoral thesis, comprehensive summary (Other academic)
Alternative title[sv]
Realtidsanalys av kolmonoxid i utandningsluften : detektion med laserspektroskopi och modellering av gasutbytet i lungorna
Abstract [en]

Breath gas analysis is a promising approach for non-invasive medical diagnostics and physiological monitoring. Real-time, breath-cycle resolved biomarker detection facilitates data interpretation and has the potential to improve the diagnostic value of breath tests as exhalation profiles carry spatiotemporal information about biomarker origin and gas exchange in the respiratory tract. This thesis presents and scrutinizes a novel methodology for the analysis of real-time breath data, where single-exhalation profiles are simulated using a pulmonary gas exchange model and least-squares fitted to measured expirograms to extract airway and alveolar contributions and diffusing capacities. The methodology is demonstrated on exhaled breath carbon monoxide (eCO), a candidate biomarker for oxidative stress and respiratory diseases. The thesis mainly covers (1) the construction of a compact optical sensor based on tunable diode laser absorption spectroscopy (TDLAS) in the mid-infrared region (4.7 μm) for selective and precise real-time detection of CO in breath and ambient air (detection limit 9 ± 5 ppb at 0.1 s), (2) the design of an advanced online breath sampling system, (3) the implementation of a trumpet model with axial diffusion (TMAD) to simulate the CO gas exchange, and (4) the application of extended eCO analysis in clinical studies to establish the healthy non-smoker baseline of the eCO parameters and to study the response to CO and wood smoke exposure. It is shown that the TMAD adequately describes the gas exchange during systemic CO elimination for different breathing patterns, and that there is no difference between eCO parameters from mouth- and nose exhalations. Expirogram shape and eCO parameters exhibit a dependence on the exhalation flow rate, but for a given breathing maneuverer, the parameters lie in a narrow range. Airway CO is close to and correlates with ambient air CO, indicating negligible airway production in the healthy population. The alveolar diffusing capacity is independent of endogenous CO, even after exposure to elevated exogenous CO, and could be used to assess lung diffusion abnormalities. Compared to CO exposure, no clear additional effect of exposure to wood smoke particles on eCO is observed. The discrimination between endogenous and exogenous CO sources remains a challenge.

Abstract [sv]

Detektion av spårgaser i utandningsluften har stor potential för icke-invasiv medicinsk diagnostik och fysiologisk övervakning. Realtid andningsgasanalys av enskilda andningscykler underlättar datatolkningen och kan förbättra det diagnostiska värdet av andningstester, eftersom utandningsprofiler bär spatiotemporal information om biomarkörens ursprung och gasutbyte i andningssystemet. Denna avhandling presenterar och granskar en ny analysmetod, där utandningsprofiler simuleras med hjälp av en matematisk modell för gasutbytet, och anpassas till uppmätta expirogram för att bestämma luftvägs- och alveolära bidrag och diffusionsförmågor. Metoden demonstreras på utandad kolmonoxid (eCO), en potentiell biomarkör för oxidativ stress och respiratoriska sjukdomar. Avhandlingen omfattar huvudsakligen (1) konstruktionen av en kompakt optisk sensor baserat på mid-infraröd diodlaserabsorptionsspektroskopi (TDLAS) vid 4.7 μm för selektiv och precis realtidsmätning av CO i utandnings- och omgivningsluften (detektionsgräns 9 ± 5 ppb vid 0.1 s), (2) design av ett avancerat system för online provtagning, (3) adaption av en matematisk lungmodell med axiell diffusion (TMAD) för simulation av CO gasutbytet, och (4) tillämpningen av utökad eCO analys i kliniska studier för att fastställa baslinjen för eCO parametrarna i friska icke-rökare, och för att studera effekten av exponering för CO och trärök. Det visas att modellen väl beskriver gasutbytet under systemiskt CO utsläpp för olika andningsmönster, och att det inte finns någon skillnad mellan eCO parametrarna från utandning via mun och näsa. Utandningsprofilerna och eCO parametrarna ändras beroende på utandningsflödet, men för ett visst andningsmönstret ligger parametrarna i ett smalt område. Koncentrationen av CO i luftvägarna ligger nära och korrelerar med CO i omgivningsluften, vilket indikerar att CO produktionen i luftvägarna är försumbart hos den friska befolkningen. Den alveolär diffusionsförmågan är oberoende av endogen CO, även efter exponering för förhöjd exogen CO, och kan möjligtvis användas för att diagnosticera en nedsatt diffusionsförmåga. Jämfört med exponering för CO observeras ingen tydlig ytterligare effekt av exponering för trärökpartiklar. Att åtskilja endogena och exogena eCO källor förbli en utmaning.

Place, publisher, year, edition, pages
Umeå: Umeå University, 2018. p. 95
Keywords
carbon monoxide (CO), pulmonary gas exchange, computational modeling, real-time breath gas analysis, single-exhalation profile, laser absorption spectroscopy, nonlinear least-squares fitting, breath sampling, baseline level, diurnal variation, healthy population, exposure study
National Category
Physiology Bioinformatics (Computational Biology) Atom and Molecular Physics and Optics Medical Laboratory and Measurements Technologies
Research subject
Physics; Physiology
Identifiers
urn:nbn:se:umu:diva-152099 (URN)978-91-7601-930-6 (ISBN)
Public defence
2018-10-19, Lilla hörsalen KBE301, KBC-huset, Umeå, 09:00 (English)
Opponent
Supervisors
Available from: 2018-09-28 Created: 2018-09-26 Last updated: 2018-10-18Bibliographically approved

Open Access in DiVA

fulltext(7931 kB)72 downloads
File information
File name FULLTEXT01.pdfFile size 7931 kBChecksum SHA-512
c3747a90d7de063b103a7ac7d84ffdb9d90bfc40275a01c00eeb6f414e434bc1ace7439a4ef38cfaa9d39090394dc1e51f6b5297219981a357a9ba0417a48bab
Type fulltextMimetype application/pdf

Other links

Publisher's full text

Search in DiVA

By author/editor
Ghorbani, RaminSchmidt, Florian M.
By organisation
Department of Applied Physics and Electronics
In the same journal
Optics Express
Medical Laboratory and Measurements TechnologiesAtom and Molecular Physics and OpticsPhysical ChemistryPhysiology

Search outside of DiVA

GoogleGoogle Scholar
Total: 72 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

doi
urn-nbn

Altmetric score

doi
urn-nbn
Total: 376 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf