Aspects of Electrical Bioimpedance Spectrum Estimation
KTH, Skolan för teknik och hälsa (STH), Medicinsk teknik, Medicinska sensorer, signaler och system.ORCID-id: 0000-0001-7807-8682
2014 (Engelska)Licentiatavhandling, sammanläggning (Övrigt vetenskapligt)
Abstract [en]

Electrical bioimpedance spectroscopy (EBIS) has been used to assess the status or composition of various types of tissue, and examples of EBIS include body composition analysis (BCA) and tissue characterisation for skin cancer detection. EBIS is a non-invasive method that has the potential to provide a large amount of information for diagnosis or monitoring purposes, such as the monitoring of pulmonary oedema, i.e., fluid accumulation in the lungs. However, in many cases, systems based on EBIS have not become generally accepted in clinical practice. Possible reasons behind the low acceptance of EBIS could involve inaccurate models; artefacts, such as those from movements; measurement errors; and estimation errors. Previous thoracic EBIS measurements aimed at pulmonary oedema have shown some uncertainties in their results, making it difficult to produce trustworthy monitoring methods. The current research hypothesis was that these uncertainties mostly originate from estimation errors. In particular, time-varying behaviours of the thorax, e.g., respiratory and cardiac activity, can cause estimation errors, which make it tricky to detect the slowly varying behaviour of this system, i.e., pulmonary oedema.

The aim of this thesis is to investigate potential sources of estimation error in transthoracic impedance spectroscopy (TIS) for pulmonary oedema detection and to propose methods to prevent or compensate for these errors.   This work is mainly focused on two aspects of impedance spectrum estimation: first, the problems associated with the delay between estimations of spectrum samples in the frequency-sweep technique and second, the influence of undersampling (a result of impedance estimation times) when estimating an EBIS spectrum. The delay between frequency sweeps can produce huge errors when analysing EBIS spectra, but its effect decreases with averaging or low-pass filtering, which is a common and simple method for monitoring the time-invariant behaviour of a system. The results show the importance of the undersampling effect as the main estimation error that can cause uncertainty in TIS measurements.  The best time for dealing with this error is during the design process, when the system can be designed to avoid this error or with the possibility to compensate for the error during analysis. A case study of monitoring pulmonary oedema is used to assess the effect of these two estimation errors. However, the results can be generalised to any case for identifying the slowly varying behaviour of physiological systems that also display higher frequency variations.  Finally, some suggestions for designing an EBIS measurement system and analysis methods to avoid or compensate for these estimation errors are discussed.

Ort, förlag, år, upplaga, sidor
Stockholm: KTH Royal Institute of Technology, 2014. , s. xi, 36
Serie
TRITA-STH : report, ISSN 1653-3836 ; 2014:4
Nyckelord [en]
Electrical Bioimpedance Spectroscopy, Thoracic Bioimpedance Spectroscopy, Sub-Nyquist Sampling, Undersampling, Aliasing in Electrical Bioimpedance, Bioimpedance Estimation Error.
Nationell ämneskategori
Medicinteknik Signalbehandling
Forskningsämne
Medicinsk teknologi; Elektro- och systemteknik
Identifikatorer
URN: urn:nbn:se:kth:diva-145643ISBN: 978-91-7595-196-6 (tryckt)OAI: oai:DiVA.org:kth-145643DiVA, id: diva2:720257
Presentation
2014-08-21, 221, Alfred Nobels Allé 10, Flemingsberg, 15:37 (Engelska)
Opponent
Handledare
Anmärkning

QC 20140604

Tillgänglig från: 2014-06-04 Skapad: 2014-05-23 Senast uppdaterad: 2014-06-04Bibliografiskt granskad
Delarbeten
1. Software tool for analysis of breathing-related errors in transthoracic electrical bioimpedance spectroscopy measurements
Öppna denna publikation i ny flik eller fönster >>Software tool for analysis of breathing-related errors in transthoracic electrical bioimpedance spectroscopy measurements
2012 (Engelska)Ingår i: Journal of Physics, Conference Series, ISSN 1742-6588, E-ISSN 1742-6596, Vol. 407, nr 1, s. 012028-Artikel i tidskrift (Refereegranskat) Published
Abstract [en]

During the last decades, Electrical Bioimpedance Spectroscopy (EBIS) has been applied in a range of different applications and mainly using the frequency sweep-technique. Traditionally the tissue under study is considered to be timeinvariant and dynamic changes of tissue activity are ignored and instead treated as a noise source. This assumption has not been adequately tested and could have a negative impact and limit the accuracy for impedance monitoring systems. In order to successfully use frequency-sweeping EBIS for monitoring time-variant systems, it is paramount to study the effect of frequency-sweep delay on Cole Model-based analysis. In this work, we present a software tool that can be used to simulate the influence of respiration activity in frequency-sweep EBIS measurements of the human thorax and analyse the effects of the different error sources. Preliminary results indicate that the deviation on the EBIS measurement might be significant at any frequency, and especially in the impedance plane. Therefore the impact on Cole-model analysis might be different depending on method applied for Cole parameter estimation.

Nyckelord
electrical bioimpedance spectroscopy, EBIS artefacts, frequency-sweep EBIS, Transthoracic Measurements, frequency-sweep spectroscopy of time-variant systems
Nationell ämneskategori
Annan medicin och hälsovetenskap
Identifikatorer
urn:nbn:se:kth:diva-118870 (URN)10.1088/1742-6596/407/1/012028 (DOI)000314985000028 ()2-s2.0-84874095138 (Scopus ID)
Konferens
1st Latin-American Conference on Bioimpedance, CLABIO 2012, 6 November 2012 through 9 November 2012, Joinville
Anmärkning

QC 20130304

Tillgänglig från: 2013-03-04 Skapad: 2013-03-04 Senast uppdaterad: 2017-12-06Bibliografiskt granskad
2. Removing respiratory artefacts from transthoracic bioimpedance spectroscopy measurements
Öppna denna publikation i ny flik eller fönster >>Removing respiratory artefacts from transthoracic bioimpedance spectroscopy measurements
Visa övriga...
2013 (Engelska)Ingår i: XV International Conference on Electrical Bio-Impedance (ICEBI) & XIV Conference on Electrical Impedance Tomography (EIT), Institute of Physics Publishing (IOPP), 2013, Vol. 434, nr 1Konferensbidrag, Publicerat paper (Refereegranskat)
Abstract [en]

Transthoracic impedance spectroscopy (TIS) measurements from wearable textile electrodes provide a tool to remotely and non-invasively monitor patient health. However, breathing and cardiac processes inevitably affect TIS measurements, since they are sensitive to changes in geometry and air or fluid volumes in the thorax. This study aimed at investigating the effect of respiration on Cole parameters extracted from TIS measurements and developing a method to suppress artifacts. TIS data were collected from 10 participants at 16 frequencies (range: 10 kHz - 1 MHz) using a textile electrode system (Philips Technologie Gmbh). Simultaneously, breathing volumes and frequency were logged using an electronic spirometer augmented with data from a breathing belt. The effect of respiration on TIS measurements was studied at paced (10 and 16 bpm) deep and shallow breathing. These measurements were repeated for each subject in three different postures (lying down, reclining and sitting). Cole parameter estimation was improved by assessing the tidal expiration point thus removing breathing artifacts. This leads to lower intra-subject variability between sessions and a need for less measurements points to accurately assess the spectra. Future work should explore algorithmic artifacts compensation models using breathing and posture or patient contextual information to improve ambulatory transthoracic impedance measurements.

Ort, förlag, år, upplaga, sidor
Institute of Physics Publishing (IOPP), 2013
Serie
Journal of Physics Conference Series, ISSN 1742-6588 ; 434:1
Nyckelord
Air, Artificial intelligence, Electric impedance, Electric impedance tomography, Parameter estimation, Patient monitoring, Textiles
Nationell ämneskategori
Annan medicinteknik
Identifikatorer
urn:nbn:se:kth:diva-136737 (URN)10.1088/1742-6596/434/1/012018 (DOI)000326512200018 ()2-s2.0-84877333426 (Scopus ID)
Konferens
15th International Conference on Electrical Bio-Impedance, ICEBI 2013 and 14th Conference on Electrical Impedance Tomography, EIT 2013; Heilbad Heiligenstadt; Germany; 22 April 2013 through 25 April 2013
Anmärkning

QC 20131209

Tillgänglig från: 2013-12-08 Skapad: 2013-12-08 Senast uppdaterad: 2016-04-18Bibliografiskt granskad
3. Electrical bioimpedance spectroscopy in time-variant systems: Is undersampling always a problem?
Öppna denna publikation i ny flik eller fönster >>Electrical bioimpedance spectroscopy in time-variant systems: Is undersampling always a problem?
2014 (Engelska)Ingår i: Journal of Electrical Bioimpedance, ISSN 1891-5469, Vol. 5, nr 1, s. 28-33Artikel i tidskrift (Refereegranskat) Published
Abstract [en]

During the last decades, Electrical Bioimpedance Spectroscopy (EBIS) has been applied mainly by using the frequency-sweep technique, across a range of many different applications. Traditionally, the tissue under study is considered to be time-invariant and dynamic changes of tissue activity are ignored by treating the changes as a noise source. A new trend in EBIS is simultaneous electrical stimulation with several frequencies, through the application of a multi-sine, rectangular or other waveform. This method can provide measurements fast enough to sample dynamic changes of different tissues, such as cardiac muscle. This high sampling rate comes at a price of reduction in SNR and the increase in complexity of devices. Although the frequency-sweep technique is often inadequate for monitoring the dynamic changes in a variant system, it can be used successfully in applications focused on the time-invariant or slowly-variant part of a system. However, in order to successfully use frequency-sweep EBIS for monitoring time-variant systems, it is paramount to consider the effects of aliasing and especially the folding of higher frequencies, on the desired frequency e.g. DC level. This paper discusses sub-Nyquist sampling of thoracic EBIS measurements and its application in the case of monitoring pulmonary oedema. It is concluded that by considering aliasing, and with proper implementation of smoothing filters, as well as by using random sampling, frequency-sweep EBIS can be used for assessing time-invariant or slowly-variant properties of time-variant biological systems, even in the presence of aliasing. In general, undersampling is not always a problem, but does always require proper consideration.

Nyckelord
Electrical bioimpedance spectroscopy, Aliasing in thoracic impedance spectroscopy, frequency-sweep EBIS, Thoracic Impedance Measurements, sub-Nyquist sampling
Nationell ämneskategori
Medicinsk apparatteknik
Forskningsämne
Medicinsk teknologi; Elektro- och systemteknik
Identifikatorer
urn:nbn:se:kth:diva-145304 (URN)10.5617/jeb.801 (DOI)2-s2.0-84902521900 (Scopus ID)
Anmärkning

QC 20140604

Tillgänglig från: 2014-05-23 Skapad: 2014-05-15 Senast uppdaterad: 2014-06-04Bibliografiskt granskad

Open Access i DiVA

Thesis(3146 kB)531 nedladdningar
Filinformation
Filnamn FULLTEXT01.pdfFilstorlek 3146 kBChecksumma SHA-512
22e0fe0388ed463d2396573812766f816257badc6192834ba09a0a5765e3c05670cfde19ef2755d95f922fabff01ba7f3963cb9e979acc2e0662d54dbbf1f0b9
Typ fulltextMimetyp application/pdf

Sök vidare i DiVA

Av författaren/redaktören
Abtahi, Farhad
Av organisationen
Medicinska sensorer, signaler och system
MedicinteknikSignalbehandling

Sök vidare utanför DiVA

GoogleGoogle Scholar
Totalt: 531 nedladdningar
Antalet nedladdningar är summan av nedladdningar för alla fulltexter. Det kan inkludera t.ex tidigare versioner som nu inte längre är tillgängliga.

isbn
urn-nbn

Altmetricpoäng

isbn
urn-nbn
Totalt: 1009 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf