Accuracy of blood flow assessment in cerebral arteries with 4D flow MRI: Evaluation with three segmentation methods
Umeå universitet, Medicinska fakulteten, Institutionen för strålningsvetenskaper.ORCID-id: 0000-0002-5911-9511
Umeå universitet, Medicinska fakulteten, Institutionen för strålningsvetenskaper.
Umeå universitet, Medicinska fakulteten, Institutionen för strålningsvetenskaper. Umeå universitet, Medicinska fakulteten, Umeå centrum för funktionell hjärnavbildning (UFBI).
Umeå universitet, Medicinska fakulteten, Institutionen för farmakologi och klinisk neurovetenskap.ORCID-id: 0000-0001-6451-1940
Vise andre og tillknytning
2019 (engelsk)Inngår i: Journal of Magnetic Resonance Imaging, ISSN 1053-1807, E-ISSN 1522-2586, Vol. 50, nr 2, s. 511-518Artikkel i tidsskrift (Fagfellevurdert) Published
Abstract [en]

Background: Accelerated 4D flow MRI allows for high‐resolution velocity measurements with whole‐brain coverage. Such scans are increasingly used to calculate flow rates of individual arteries in the vascular tree, but detailed information about the accuracy and precision in relation to different postprocessing options is lacking.

Purpose: To evaluate and optimize three proposed segmentation methods and determine the accuracy of in vivo 4D flow MRI blood flow rate assessments in major cerebral arteries, with high‐resolution 2D PCMRI as a reference.

Study Type: Prospective.

Subjects: Thirty‐five subjects (20 women, 79 ± 5 years, range 70–91 years).

Field Strength/Sequence: 4D flow MRI with PC‐VIPR and 2D PCMRI acquired with a 3 T scanner.

Assessment: We compared blood flow rates measured with 4D flow MRI, to the reference, in nine main cerebral arteries. Lumen segmentation in the 4D flow MRI was performed with k‐means clustering using four different input datasets, and with two types of thresholding methods. The threshold was defined as a percentage of the maximum intensity value in the complex difference image. Local and global thresholding approaches were used, with evaluated thresholds from 6–26%.

Statistical Tests: Paired t‐test, F‐test, linear correlation (P < 0.05 was considered significant) along with intraclass correlation (ICC).

Results: With the thresholding methods, the lowest average flow difference was obtained for 20% local (0.02 ± 15.0 ml/min, ICC = 0.97, n = 310) or 10% global (0.08 ± 17.3 ml/min, ICC = 0.97, n = 310) thresholding with a significant lower standard deviation for local (F‐test, P = 0.01). For all clustering methods, we found a large systematic underestimation of flow compared with 2D PCMRI (16.1–22.3 ml/min).

Data Conclusion: A locally adapted threshold value gives a more stable result compared with a globally fixed threshold. 4D flow with the proposed segmentation method has the potential to become a useful reliable clinical tool for assessment of blood flow in the major cerebral arteries.

Level of Evidence: 2

Technical Efficacy: Stage 2

sted, utgiver, år, opplag, sider
2019. Vol. 50, nr 2, s. 511-518
Emneord [en]
magnetic resonance imaging, cerebral arteries, phase‐contrast MRI, 4D flow MRI, cerebral blood flow, Circle of Willis
HSV kategori
Identifikatorer
URN: urn:nbn:se:umu:diva-147255DOI: 10.1002/jmri.26641ISI: 000475681600017PubMedID: 30637846Scopus ID: 2-s2.0-85059966858OAI: oai:DiVA.org:umu-147255DiVA, id: diva2:1202720
Forskningsfinansiär
Swedish Research Council, 2015–05616; 2017-04949Swedish Heart Lung Foundation, 20140592
Merknad

Originally included in thesis in manuscript form with title "Blood flow assessment in cerebral arteries with 4D flow MRI, concordance with 2D PCMRI"

Tilgjengelig fra: 2018-04-30 Laget: 2018-04-30 Sist oppdatert: 2019-08-12bibliografisk kontrollert
Inngår i avhandling
1. Blood flow assessment in cerebral arteries with 4D flow magnetic resonance imaging: an automatic atlas-based approach
Åpne denne publikasjonen i ny fane eller vindu >>Blood flow assessment in cerebral arteries with 4D flow magnetic resonance imaging: an automatic atlas-based approach
2018 (engelsk)Doktoravhandling, med artikler (Annet vitenskapelig)
Alternativ tittel[sv]
Blodflödesmätning i cerebrala artärer med 4D flödes magnetresonanstomografi : en automatisk atlasbaserad metod
Abstract [en]

Background: Disturbed blood flow to the brain has been associated with several neurological diseases, from stroke and vascular diseases to Alzheimer’s and cognitive decline. To determine the cerebral arterial blood flow distribution, measurements are needed in both distal and proximal arteries.

4D flow MRI makes it possible to obtain blood flow velocities from a volume covering the entire brain in one single scan. This facilitates more extensive flow investigations, since flow rate assessment in specific arteries can be done during post-processing. The flow rate assessment is still rather laborious and time consuming, especially if the number of arteries of interest is high. In addition, the quality of the measurements relies heavily on the expertise of the investigator.

The aim of this thesis was to develop and evaluate an automatic post-processing tool for 4D flow MRI that identifies the main cerebral arteries and calculates their blood flow rate with minimal manual input. Atlas-based labeling of brain tissue is common in toolboxes for analysis of neuroimaging-data, and we hypothesized that a similar approach would be suitable for arterial labeling. We also wanted to investigate how to best separate the arterial lumen from background for calculation of blood flow.

Methods: An automatic atlas-based arterial identification method (AAIM) for flow assessment was developed. With atlas-based labeling, voxels are labeled based on their spatial location in MNI-space, a stereotactic coordinate system commonly used for neuroimaging analysis. To evaluate the feasibility of this approach, a probabilistic atlas was created from a set of angiographic images derived from 4D flow MRI. Included arteries were the anterior (ACA), middle (MCA) and posterior (PCA) cerebral arteries, as well as the internal carotid (ICA), vertebral (VA), basilar (BA) and posterior communicating (PCoA) arteries. To identify the arteries in an angiographic image, a vascular skeleton where each branch represented an arterial segment was extracted and labeled according to the atlas. Labeling accuracy of the AAIM was evaluated by visual inspection.

Next, the labeling method was adapted for flow measurements by pre-defining desired regions within the atlas. Automatic flow measurements were then compared to measurements at manually identified locations. During the development process, arterial identification was evaluated on four patient cohorts, with and without vascular disease. Finally, three methods for flow quantification using 4D flow MRI: k-means clustering; global thresholding; and local thresholding, were evaluated against a standard reference method.

Results: The labeling accuracy on group level was between 96% and 87% for all studies, and close to 100% for ICA and BA. Short arteries (PCoA) and arteries with large individual anatomical variation (VA) were the most challenging. Blood flow measurements at automatically identified locations were highly correlated (r=0.99) with manually positioned measurements, and difference in mean flow was negligible.

Both global and local thresholding out-performed k-means clustering, since the threshold value could be optimized to produce a mean difference of zero compared to reference. The local thresholding had the best concordance with the reference method (p=0.009, F-test) and was the only method that did not have a significant correlation between flow difference and flow rate. In summary, with a local threshold of 20%, ICC was 0.97 and the flow rate difference was -0.04 ± 15.1 ml/min, n=308.

Conclusion: This thesis work demonstrated that atlas-based labeling was suitable for identification of cerebral arteries, enabling automated processing and flow assessment in 4D flow MRI. Furthermore, the proposed flow rate quantification algorithm reduced some of the most important shortcomings associated with previous methods. This new platform for automatic 4D flow MRI data analysis fills a gap needed for efficient in vivo investigations of arterial blood flow distribution to the entire vascular tree of the brain, and should have important applications to practical use in neurological diseases.

sted, utgiver, år, opplag, sider
Umeå: Umeå universitet, 2018. s. 60
Serie
Umeå University medical dissertations, ISSN 0346-6612 ; 1965
Emneord
Circle of Willis, 4D flow MRI, Cerebral arteries, Vascular disease, Stroke, Automatic labeling, Probabilistic atlas, Cerebral blood flow, Neuroimaging, Magnetic Resonance Imaging
HSV kategori
Identifikatorer
urn:nbn:se:umu:diva-147256 (URN)978-91-7601-889-7 (ISBN)
Disputas
2018-05-25, Betula, NUS, Umeå, 13:00 (svensk)
Opponent
Veileder
Forskningsfinansiär
Swedish Research Council, 2015–05616Swedish Heart Lung Foundation, 20110383Swedish Heart Lung Foundation, 20140592The Swedish Brain Foundation
Tilgjengelig fra: 2018-05-04 Laget: 2018-04-30 Sist oppdatert: 2018-06-09bibliografisk kontrollert

Open Access i DiVA

fulltext(529 kB)40 nedlastinger
Filinformasjon
Fil FULLTEXT02.pdfFilstørrelse 529 kBChecksum SHA-512
8402feddf5b455fb15650e38609bd40f3b769f3e48812bf2014eda3c3789fc86098fc7b115f979679991110c43c0e2288379335d97bc76b9f45a783168b48635
Type fulltextMimetype application/pdf

Andre lenker

Forlagets fulltekstPubMedScopus

Søk i DiVA

Av forfatter/redaktør
Dunås, ToraHolmgren, MadeleneWåhlin, AndersMalm, JanEklund, Anders
Av organisasjonen
I samme tidsskrift
Journal of Magnetic Resonance Imaging

Søk utenfor DiVA

GoogleGoogle Scholar
Totalt: 81 nedlastinger
Antall nedlastinger er summen av alle nedlastinger av alle fulltekster. Det kan for eksempel være tidligere versjoner som er ikke lenger tilgjengelige

doi
pubmed
urn-nbn

Altmetric

doi
pubmed
urn-nbn
Totalt: 527 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf